Devicemaker issues software fix for glucose tests: 3 notes

Nova Biomedical issued an urgent software correction for its StatStrip Glucose and Glucose/Ketone Hospital Meter Systems after it found a defect that could transmit incorrect glucose and ketone results to hospital data management systems.

Advertisement

Here are three notes: 

  1. The company on Nov. 20 alerted customers to the issue, which was identified as a potential risk to patient safety, according to a Jan. 30 FDA news release. The agency has classified this recall as the most serious type. 
  2. The issue arises when the meter’s “review results” screen is accessed while the device is transmitting current test results. This could lead to inaccurate historical test results being recorded in the hospital’s medical record system. 

If left unaddressed, this could result in misdiagnoses such as incorrect treatment for hypoglycemia, hyperglycemia or ketoacidosis particularly in high-risk populations like neonates and intensive care patients.

3.  No injuries or deaths have been reported. Healthcare providers have been instructed to distribute the correction notice and report any instances of incorrect data      transmission to Nova Biomedical’s technical support team.

Becker’s 3rd Annual Fall Chief Pharmacy Officer Summit is the top tier event that the people leading the future of pharmacy convene. This 3-day event is built specifically for Chief Pharmacy Officers and senior pharmacy leaders. The sessions examine the workforce, financial and clinical priorities shaping pharmacy today. Becker’s is where candid discussions flourish and new ideas thrive. Becker's 3rd Annual Fall Chief Pharmacy Officer Summit