Certain Samaritan 300/300P PAD devices have been found to intermittently turn on and off, which may eventually deplete the battery. In addition and separately, certain Samaritan 300/300P PAD devices containing early versions of the battery management software may misinterpret a temporary drop in battery voltage as signaling a low battery, resulting in the device turning itself off. A device experiencing either condition could be unable to deliver therapy during a cardiac event.
Samaritan 300/300P PAD devices with the following serial numbers are affected by this recall:
- 0400000501 to 0700032917
- 08A00035000 to 10A0070753
- 10C00200000 to 10C00210106
Related Articles on Recalls:
Mindray Anesthesia Delivery System Voluntarily Recalled
ASP Recalls All Lots of Sterrad Cyclesure 24 Biological Indicators
Baxter Buretrol Solution Sets Recalled
At the Becker's 11th Annual IT + Revenue Cycle Conference: The Future of AI & Digital Health, taking place September 14–17 in Chicago, healthcare executives and digital leaders from across the country will come together to explore how AI, interoperability, cybersecurity, and revenue cycle innovation are transforming care delivery, strengthening financial performance, and driving the next era of digital health. Apply for complimentary registration now.